Mi180 pH / ORP / Conductivity / TDS / NaCl / Temperature Laboratory Bench Meter

Mi180 measures 6 different parameters: pH, ORP, EC, TDS (Total Dissolved Solids), percentage of NaCl and temperature in a variety of ranges.
pH calibration can be performed in 3 points selectable between 7 memorized buffers, to provide a very accurate calibration curve even when testing different samples, where very wide differences in pH can be found.

The auto-ranging feature for EC and TDS measurements automatically sets the resolution suitable to the tested sample.
All measurements can be temperature compensated at 20 or 25°C and the compensation coefficient is selectable by the user.

The automatic temperature compensation can also be disabled for measuring the actual conductivity value. The stability indicator on the LCD ensures accuracy. Conductivity readings are performed with the 4-ring probe supplied with the meter.
The GLP feature allows users to store and recall data on system status. PC compatible through an RS232 port or USB.


 

SPECIFICATIONS
pH Range -2.00 to 16.00 pH; -2.000 to 16.000 pH
mV Range ±699.9 mV; ±2000 mV
EC Range 0.00 to 29.99 µS/cm; 30.0 to 299.9 µS/cm; 300 to 2999 µS/cm; 3.00 to 29.99 mS/cm; 30.0 to 200.0 mS/cm; up to 500.0 mS/cm uncompensed EC*
TDS Range 0.0 to 14.99 mg/l (ppm); 15.0 to 149.9 mg/l (ppm); 150 to 1499 mg/l (ppm); 1.50 to 14.99 g/l (ppt); 15.0 to 100.0 g/l (ppt); up to 400.0 g/l actual TDS* (with 0.80 factor)
NaCl Range 0.0 to 400.0%
Temp Range -20.0 to 120.0°C / -4.0 to 248.0°F
pH Resolution 0.01 pH; 0.001 pH
mV Resolution 0.1 mV; 1 mV
EC Resolution 0.01 µS/cm; 0.1 µS/cm; 1 µS/cm; 0.01 mS/cm; 0.1 mS/cm;
TDS Resolution 0.01 mg/l; 0.1 mg/l; 1.0 mg/l; 0.01 g/l; 0.1 g/l
NaCl Resolution 0.1%
Temp Resolution 0.1°C / 0.1°F
pH Accuracy ±0.01 pH; ±0.002 pH
mV Accuracy ±0.2 mV; ±1 mV
EC Accuracy ±1% of reading ±(0.05 µS/cm or 1 digit)
TDS Accuracy ±1% of reading ±(0.03 ppm or 1 digit)
NaCl Accuracy ±1% of reading
Temp Accuracy ±0.4°C / ±0.8°F
Rel mV Offset ±2000 mV
pH Calibration 1, 2 or 3 points calibration, with 7 memorized buffers (pH 1.68, 4.01, 6.86, 7.01, 9.18, 10.01 and 12.45)
EC Calibration 1 point slope calibration with 6 memorized solutions: (84 µS/cm, 1413 µS/cm, 5.00 mS/cm, 12.88 µS/cm, 80.0 µS/cm, 111.8 mS/cm)
NaCl Calibration 1 point
Temp Calibration 2 point, at 0 and 50°C / 32 and 122°F
Temperature Compensation automatic or manual, from -20.0 to 120.0°C / -4.0 to 248.0°FF
Temperature Coefficient selectable from 0.00 to 6.00%/°C (EC and TDS only)
pH Electrodes & Temp Probe MA814DB/1 & MA917B/1 & MA831R (included)
TDS Factor 0.40 to 0.80 (default value is 0.50)
Log on demand up to 50 samples on each range (pH, mV, EC, TDS, NaCl)
GLP last pH, EC, NaCl calibration data
PC Interface RS232 / USB Opto-isolated
Environment 0 to 50°C / 32 to 122°F; max RH 95%
Power supply 12 VDC power adapter (included)
Dimensions 230 x 160 x 95 mm
Weight 0.9 kg

 