The Nano-100 is a perfect instrument for micro-volume UV-VIS spetrophotometer. It can accurately measure DNA, RNA and Oligonucleotide concentrations, protein assays etc.. The Nano-100 utilises CCD detector, a light source for long lifetime Xenon lamp, together with simple software. Simply pipette 0.5~2μl sample onto the pedestal, and the measurement can be completed within 5 seconds.
Adding sample measuring sample quick and easy cleaning
Features
1. Turn on and instantly measure without lamp warm up time
2. Full range of wavelength (200-800nm) with scan capability from 200-800nm within 5sec.
3. User-friendly software with free software updates
4. Simply pipette a sample onto the pedestal, measure, wipe the pedestal
5. Direct micro-volume measurement with only 0.5-2ul sample required to accurately determinate nucleic acids and protein concentrations
6. Long life Xenon flash lamp, 10 flashes, up to 10 years
7. No cell or cuvette and expensive consumable required
8. Detects Nucleic Acids up to 4500ng/ul (dsDNA)
9. 3864-Element linear silicon CCD array
Model | Nano-100 |
Wavelength Range | 200~800nm |
Sample Size | 0.5~2.0ul |
Path Length | 0.2mm (For high concentration measwrement) 1.0mm (For ordinary) |
Light Source | Xenon flash lamp |
Detector Type | 3864-element linear silicon CCD array |
Wavelength Accuracy | 1nm |
Spectral Resolution | ≤3nm (FWHM at Hg 546nm) |
Absorbance Precision | 0.003Abs |
Absorbance Accuracy | 1% (7.332Abs at 260nm) |
Absorbance Range | 0.02 - 90A |
Detection Range | 2~4500ng/ul (dsDNA) |
Measurement Time | <8s |
Data Output | Connect PC |
Dimensions (W x D x H)mm | 200 x 250 x 166 |
Weight | 2.6kg |
Sample Pedestal Material | Aluminum alloy and Quartz fiber |
Operating Voltage | 24VDC |
Operating Power Consumption | 20W |
Standby Power Consumption | 5W |
Software Compatibility | Win XP,Windows 7, Windows 8 |